An MRI scanner is unlike any other equipment purchase a Dallas practice will make. The magnet is always on, the room has to be built around it, and the project involves riggers, shielding contractors, cooling systems and safety planning long before the first scan. A refurbished system can lower the purchase price considerably, but the site work and running costs are much the same either way. This guide sets out what is involved so you can judge a quote and plan a realistic timeline. For other imaging and clinical equipment, see our new and refurbished medical equipment for sale.
First decision: what kind of MRI
| Type | Typical use | Site impact |
|---|---|---|
| High-field closed bore (1.5T) | General outpatient and hospital imaging | Full shielded room, cooling, helium management |
| 3T closed bore | Neuro, MSK and advanced imaging | Larger fringe field, more shielding and planning |
| Open or low-field | Claustrophobic or larger patients, some orthopedic work | Often lighter siting requirements; lower image quality for some studies |
| Extremity MRI | Knees, ankles, wrists, elbows in orthopedic practices | Compact room; limited to extremities |
| Mobile MRI trailer | Shared or temporary service | Pad, power and connections rather than a built room |
Start from your study mix and referral sources. An orthopedic group scanning mostly knees and shoulders has different needs from a multi-specialty imaging center.
Siting: the room is built around the magnet
- RF shielding: the scan room is enclosed in a radiofrequency shield (a Faraday cage) to keep outside signals out of the images.
- Magnetic fringe field: the field extends beyond the magnet. Its boundary determines where people with pacemakers and other implants can safely go, and how close other equipment can be.
- Floor loading: magnets are extremely heavy; the floor must be engineered for the load.
- Rigging path: the magnet usually enters through an opening in a wall or roof, planned in advance.
- Cooling: a chiller and cryogen system keep the magnet and electronics at temperature, with dedicated power.
- Quench pipe: superconducting magnets need a vent to carry helium safely outside if the magnet quenches.
- Vibration and nearby steel: elevators, parking and heavy traffic nearby can affect image quality.
Safety zones
MRI suites are usually organized into four safety zones based on the American College of Radiology approach, as summarized by MRI Questions: Zone I is public space, Zone II is where patients are screened, Zone III is restricted to screened people near the magnet room, and Zone IV is the magnet room itself. The layout of reception, changing rooms, the control room and doors should follow these zones, so screening happens before anyone can reach the magnet. Ferromagnetic objects such as oxygen tanks, carts and tools are among the most serious hazards, so plan MR-safe or MR-conditional equipment for the suite.
Accreditation and billing
Medicare requires suppliers of the technical component of advanced diagnostic imaging, including MRI, in non-hospital settings to be accredited. CMS approved the American College of Radiology, the Intersocietal Accreditation Commission and The Joint Commission as accrediting organizations. Accreditation reviews image quality, staff qualifications, safety and equipment quality control, so plan for it from the start.
Refurbished MRI: what to verify
- Magnet field strength, age, and helium level and consumption history
- Whether the magnet will be shipped at field ("ramped") or ramped down, and who handles ramping at your site
- Coils included for the body regions you will scan
- Software version, available sequences and remaining manufacturer or independent service support
- Chiller and cold head condition and service history
- De-installation, rigging, shipping and re-installation, and who is responsible for each step
A typical project sequence
- Define the study mix and choose the type of MRI.
- Get the vendor's site planning guide and a site survey.
- Design the suite around the safety zones, with shielding and structural engineering.
- Build, install the RF shield and test it.
- Rig the magnet in, install and ramp to field.
- Run acceptance testing, train staff and start accreditation.
Costs beyond the scanner
Budget for construction and shielding, the chiller and power, rigging, installation and ramping, service contracts, helium management, accreditation, staffing and MR-safe ancillary equipment such as monitors, stretchers and injectors. For many projects these costs are a large share of the total, whether the scanner is new or refurbished.
Equipping the rest of the suite
Patient monitoring, contrast injection and transport inside Zone IV must be MR-safe or MR-conditional for your field strength. Outside the magnet room, the suite needs standard equipment such as stretchers, crash carts and patient monitors. Our patient monitor buying guide covers monitors for prep and recovery areas.
How we can help
We work with Dallas–Fort Worth practices on equipment planning and supply around imaging projects, from our Garland headquarters. Ask us whether we can source a specific MRI system for your project, and we will tell you plainly what we can and cannot supply. Our medical equipment planning and consulting team can help coordinate the equipment list around your architect and imaging vendor.
Once a scanner is installed, keeping it running is its own job. Our guide to MRI machine repair in Dallas covers common faults, what to do when the system goes down and preventive maintenance.
To talk through an MRI project, call (469) 767-8853 or send us a request with your study mix and site.