Most of the work in opening a stress lab happens outside the purchase order. The stress system itself arrives in a few crates, but the room, the electrical supply, the emergency equipment and the people running the test decide whether the lab is safe and efficient. This guide is for Austin cardiology, primary care and multi-specialty practices adding exercise testing, and covers what to prepare before the system arrives. Stress systems are part of our new and refurbished medical equipment for sale.
The room
- Space around the treadmill: staff need clear access to both sides and the rear of the treadmill, and room to lay a patient flat and start resuscitation if needed.
- Stretcher access: a stretcher must be able to reach the treadmill and leave the room without moving furniture.
- Ceiling height: a tall patient on a raised treadmill at full grade needs headroom; check the treadmill's platform height against your ceiling.
- Floor: level and strong enough for the treadmill and patient; avoid placing it over a raised access floor without checking the load.
- Climate: a cool, well-ventilated room helps patients reach target heart rate comfortably and keeps electronics stable.
- Privacy and changing: a changing area and a way to cover the patient during electrode placement.
- Sink: for skin preparation and cleaning.
Electrical supply
Commercial treadmills draw a large current when they start and climb. Many manufacturers specify a dedicated circuit of a particular voltage and amperage. Give your electrician the treadmill's installation requirements before construction, and plug the treadmill directly into its outlet rather than through an extension cord or power strip. The stress PC, printer and BP unit usually share a separate outlet.
Emergency equipment
Exercise testing is generally safe, but serious events can occur. The AHA exercise testing standards (Fletcher et al., Circulation 2013) say a defibrillator and appropriate medications should be immediately available and that the person conducting the test must be trained in advanced cardiopulmonary resuscitation. In practice, that means:
- A defibrillator in the room, checked daily. Our defibrillator buying guide compares AEDs and manual units.
- A crash cart with emergency medications, airway equipment and suction, as your policy specifies
- Oxygen
- A way to call for help immediately, such as a wall phone or emergency button that every staff member knows
- A stretcher or exam table close to the treadmill
The treadmill's own safety features
- Front and side handrails for stability
- An emergency stop button that staff can reach quickly, and a safety lanyard or stop where the model provides one
- Smooth, gradual start and stop
- Accurate speed and grade, which the AHA standards say must be calibrated
Staffing and supervision
The AHA standards describe testing by well-trained staff, with a physician generally immediately available, while allowing trained non-physicians to supervise tests in certain lower-risk groups. Your practice's policy should state who may supervise which patients, the indications and contraindications you follow, and the criteria for stopping a test. Train staff on electrode placement and skin preparation too; poor prep is the most common cause of a noisy, hard-to-read stress ECG.
Maintenance after opening
A stress lab depends on a treadmill that runs smoothly under load every day. Belts stretch and wear, decks need lubrication or replacement on the manufacturer's schedule, and motors and elevation drives need periodic checks. The ECG module, patient cable and lead wires take constant handling and should be inspected regularly, and the stress PC needs its software and operating system kept supported. Schedule preventive maintenance at least as often as the manufacturer recommends, and keep a simple log of belt adjustments, deck service and calibration checks so trends are easy to spot before a failure stops testing.
Workflow and records
- Pre-test screening checklist and consent
- Protocol selection guidance for staff
- Recovery area or chairs for post-test monitoring
- How reports reach the reading physician and the chart
- A log of daily defibrillator checks and treadmill maintenance
Choosing the system
If you are still comparing stress systems, our stress test system buying guide explains the components and families. Refurbished systems from us come with a 90-day warranty and a test record covering the treadmill, ECG module and software.
Opening-day checklist
- Treadmill installed, leveled and on its dedicated circuit
- Speed, grade and emergency stop tested
- ECG module verified with a simulator and a staff test
- Defibrillator, crash cart and oxygen in place and checked
- Protocols, report templates and output tested end to end
- Staff trained on the system and emergency response
Delivery and service in Austin
Austin is about 195 miles from our Garland headquarters. We confirm delivery and installation dates when you order, and service visits are scheduled in advance. Service requests receive a response in under 8 business hours, Monday through Friday.
To plan a stress lab in Austin, call (469) 767-8853 or send us a request with your room layout and expected test volume.