An X-ray machine is the one purchase in most Fort Worth clinics that comes with its own state paperwork. The equipment has to be registered, tested after installation and on a schedule, operated by credentialed staff and overseen by a named radiation safety officer. None of this is difficult once you know the steps, but missing one can delay opening or create a problem at inspection. This guide sets out the steps in the order they happen. For the equipment itself, see our X-ray machine repair, sales and rental page for the Dallas–Fort Worth area.
This is a practical summary, not legal advice. The Texas Department of State Health Services (DSHS) rules in 25 TAC Chapter 289 are the authority, and rules can change, so confirm details with DSHS or your medical physicist.

Step 1: Before you buy, talk to a medical physicist
A licensed medical physicist is part of almost every X-ray compliance step. DSHS directs shielding questions to the physicist associated with your facility, and the physicist will review your room layout, walls, doors and the areas next to the room. Bringing one in before you choose the room avoids rebuilding walls later.
Step 2: Name a radiation safety officer
Each DSHS registration has one radiation safety officer (RSO), who is responsible for making sure staff are trained and exposures are kept as low as reasonably achievable. According to the DSHS X-ray FAQ, a change of RSO must be reported in writing within 30 days. In a small clinic the RSO is often the physician owner or practice manager.
Step 3: Install the machine
Installation covers mounting, electrical connection, calibration, detector setup and network connection. Ask whoever installs it how they handle the equipment performance evaluation that follows (Step 4), and keep their installation records in your radiation file.
Step 4: Equipment performance evaluation within 30 days
An equipment performance evaluation (EPE) measures the machine's radiation output, timer, beam quality, collimation and other parameters. Under the DSHS EPE guidance for service companies, medical X-ray machines need an EPE within 30 days after initial installation, after reinstallation, and after repair of a component that affects radiation output, such as the tube, timer or power supply. For medical units, the EPE is performed by or under the supervision of a licensed medical physicist.
Step 5: Register with DSHS within 30 days of first use
For most fixed medical X-ray equipment, DSHS says a facility must apply for its certificate of registration within 30 days of beginning use. Some uses, including mobile X-ray services, need the certificate before use begins. Registration fees are due at the start and every two years. Adding equipment or a new site under the same legal entity uses a separate form, and the business name must match your filing with the Texas Secretary of State.
Step 6: Make sure operators are credentialed
Texas regulates who can take X-rays. Depending on the role and the procedures, staff may need certification as a medical radiologic technologist or registration in a limited category through the Texas Medical Board. Check credentials before the machine goes into use, and keep copies on file.
Step 7: Keep up the testing schedule
After the first EPE, testing continues at regular intervals. A summary of the rule by Maven Imaging puts the interval for medical radiographic units at every two years and for fluoroscopic units, including C-arms, at least annually. Put the next due date in your calendar the day you receive each report.
Step 8: Keep the radiation file in order
Inspectors expect to see the registration certificate, the RSO designation, EPE reports, shielding documentation, staff credentials, training records and any service records that affect radiation output. Keeping them in one binder or folder per machine makes inspections straightforward.
Common mistakes
- Choosing the room before the physicist has reviewed shielding.
- Treating the EPE as optional after a tube or generator repair.
- Letting staff take images before their credentials are confirmed.
- Losing track of the next EPE due date when a staff member who managed it leaves.
Changes and disposal
- Moving the machine: reinstallation triggers a new EPE within 30 days.
- Change of ownership or name: contact DSHS; a new owner may need its own registration.
- Retiring a machine: notify DSHS that the unit is out of service and follow disposal rules. Our medical equipment disposition service provides documented disposal, with records for your registration file.
Where we fit in
We help Fort Worth facilities choose equipment that fits the room and the exams, coordinate installation timing with your physicist, and provide maintenance and repair once the machine is in service. Service requests receive a response in under 8 business hours, Monday through Friday, with same-day on-site support across Dallas–Fort Worth. If you are still choosing a system, our guide to X-ray system types is the place to start.
To plan an X-ray purchase with the compliance steps in mind, call (469) 767-8853 or send us a request.