A refurbished hospital bed can cost far less than a new one and give years of reliable service, but only if the parts that wear have been checked and replaced. For San Antonio hospitals, nursing facilities, home health agencies and families, this checklist covers what should be tested on a used or refurbished bed before it reaches a patient, and the red flags that mean a bed is not ready. Beds are part of our new and refurbished medical equipment for sale.

The checklist
1. Frame and deck
- No cracked welds, bent sections or loose fasteners
- Deck sections move freely without binding
- Head and foot boards secure
2. Motors and actuators
- Every motion (head, knee, height, Trendelenburg and chair position, where fitted) runs smoothly through its full range
- No grinding, clicking or hesitation under a patient's weight
- No drift: sections hold position with the power off
3. Controls
- Hand pendant, side-rail controls and footboard panel all respond correctly
- Lockout functions work, so staff can disable controls a patient should not use
- Cords free of cuts and crushed sections
4. CPR release
The emergency CPR release must lower the head section quickly and flatten the bed. It is checked every time a bed is refurbished.
5. Side rails
- Rails raise, lower and latch securely
- No cracks, broken latches or loose mounting
- Rails match the bed model, with no mixed parts that change gap sizes
6. Brakes and casters
- Brakes lock all casters firmly
- Steer mode, where fitted, works
- Casters roll freely with no flat spots
7. Battery
Many electric beds have a backup battery so they can be lowered or moved during a power loss. Check that it holds a charge and replace it if it doesn't.
8. Scale and bed-exit alarm
Where the bed has an integrated scale, its accuracy is checked with known test weights. Bed-exit alarms are tested in each zone. Our guide to hospital bed scales explains scale checks in more detail.
9. Mattress
- Correct mattress for the bed, sized to the deck
- Cover intact with no cracks, tears or stains that soak through
- Foam not compressed or bottoming out
- Powered mattresses inflate and alarm correctly
10. Electrical safety and recalls
- Electrical safety test results recorded
- Model checked in the FDA medical device recall database, with any required corrections completed
11. Cleaning
The bed should be cleaned and disinfected, including rails, controls and under-deck areas, before delivery.
A ten-minute test you can do yourself
Even with a test report, it is worth spending a few minutes with the bed before you accept it. Lie on it, or have a colleague lie on it, and run every motion with someone on the deck, because many faults only appear under load. Raise the bed to full height and leave it for a few minutes to see whether it drifts. Lower and latch each rail and push on it firmly. Lock the brakes and try to push the bed. Pull the CPR release and confirm the head section drops smoothly. Finally, unplug the bed and confirm the battery can still lower it. If any of these feel wrong, ask for the bed to be checked again before it goes into use.
Keeping a refurbished bed in service
After delivery, put the bed on a preventive maintenance schedule, usually based on the manufacturer's recommendations. Each visit should check the motions, controls, rails, brakes, battery, scale and mattress, and record the results. Beds that are maintained this way typically give many more years of service than beds that are only repaired when they fail.
Red flags
- A bed sold "working" with no list of what was tested
- Rails or mattress from a different model
- Sections that drift down with the power off
- A CPR release that sticks
- Torn or cracked mattress covers
- No warranty in writing
What the paperwork should show
Ask for the model and serial number, the checks performed and results, parts replaced (such as actuators, pendants, batteries or casters), electrical safety results and the warranty terms. Reconditioned beds from us carry a 90-day warranty. For the documents to keep with any pre-owned equipment, see our pre-owned equipment paperwork guide.
Which beds make good refurbished buys
Well-known long-term care and acute care beds from large manufacturers are the strongest refurbished candidates, because parts, mattresses and rails remain available for years. Before buying an older or less common model, confirm actuators, pendants and rails can still be sourced.
Delivery and service in San Antonio
San Antonio is about 275 miles from our Garland headquarters. We confirm delivery dates when you order, and setup and service visits are scheduled in advance. Service requests receive a response in under 8 business hours, Monday through Friday. For repairs across Texas, see our hospital bed repair, rental and sales page.
To ask about refurbished beds and their test records, call (469) 767-8853 or send us a request.
Related guide: Hospital Bed for Sale in College Station: How to Read a Spec Sheet
Shop hospital beds and other patient-care equipment online at our store, Mr. Bedmed.
Need a bed for a short stay at home or a census surge? See our hospital bed rental options across Texas.
On a budget? See our used hospital beds for sale, reconditioned and tested.